Specifications
xBud is a multi-channel analog front-end (AFE) that conditions EEG, ECG, EMG, and EOG signals and feeds a data-acquisition device (DAQ) for digitisation. This page lists the platform specifications. xBud is DAQ-agnostic; the NI mioDAQ (USB-6451) values below are the reference example.
Applies to: xBud hardware Rev A, with BXI Studio 1.0.0 or later.
Note: xBud is for research, education, and prototyping only. It is not a medical device and is not intended for diagnosis, treatment, or monitoring. See Power and safety.
Note: Specifications marked To be confirmed are pending final hardware characterisation. All electrical figures carry their test conditions where measured.
Key features
Multi-channel EXG amplifier: 4, 8, 12, or 16 channels, plus dedicated REF, AGND, and DRL
Multimodal acquisition of EEG, ECG, EMG, and EOG from one front-end
Dry Ag/AgCl electrode support across four subject interfaces: EEG cap, headband, armband, and chest-strap
Fixed front-end conditioning: 100× gain, 300 Hz low-pass, 0.5 Hz AC-coupled high-pass
Active noise cancellation with a driven-right-leg (DRL) circuit
DAQ-agnostic: tested with NI USB-600x, NI mioDAQ, and NI cDAQ
Digitisation resolution and sampling rate set by the DAQ you pair with
Two digital marker lines via a two-button marker box
System architecture
Subject xBud (Stage 1) DAQ (Stage 2) Host PC
┌──────────┐ EXG ┌──────────────────────┐ cond.┌────────────────────┐ ┌────────────┐
│ dry EXG ├───────►│ instrumentation amp ├─────►│ analog inputs │ │ │
│electrodes│ µV–mV │ + fixed gain (100×) │ ±V │ ADC, 12–24-bit, │USB│ BXI Studio │
│ REF / DRL│ │ + fixed 300 Hz LPF │ │ rate set by the DAQ├──►│ │
└──────────┘ │ + active noise canc. │ │ │ └────────────┘
┌──────────┐ cable │ + DRL drive │ │ digital inputs │
│marker box├───────►│ 2 digital marker outs├─────►│ (marker lines) │
│ 2 buttons│ └──────────────────────┘ └────────────────────┘
└──────────┘ xBud analog + marker outputs → DAQ terminals (jumper-wired)
DAQ examples: NI USB-600x, NI mioDAQ (USB-6451), NI cDAQCapabilities and applications
Multimodal acquisition (EEG/ECG/EMG/EOG)
Biomedical instrumentation teaching
Fixed, characterised gain and bandwidth
Biosignal and digital signal processing labs
Active noise cancellation and DRL
Brain–computer interface (BCI) prototyping
Stimulus marker and triggering
Cognitive and attention studies
Up to 24-bit digitisation, DAQ-dependent
Evoked-response studies (SSVEP, AEP, somatosensory)
Real-time streaming pipelines
Surface-EMG gesture classification
Electrical: xBud analog front-end
EXG channels
4 / 8 / 12 / 16, + REF + AGND + DRL
Variant-dependent
Electrode type
Dry Ag/AgCl
Can be used with gel
Gain
100×
Fixed
Low-pass cutoff
300 Hz
Fixed
High-pass coupling
AC-coupled, 0.5 Hz
Noise floor
< 10 µV
Measured over 0.5–45 Hz
Active noise cancellation
DRL-based common-mode suppression
Minimum detectable signal (EEG)
~10 µV
Comparable to the noise floor; averaging or spectral methods recover smaller signals
Input impedance
> 500 MΩ
At DC and in-band
CMRR
> 60 dB
At 50/60 Hz
Channel crosstalk
< -80 dB
Weight
300 g
Supported DAQs
xBud is DAQ-agnostic, it produces channel output and GND for each channel. Internally, all GND pins are connected. The same front-end (4/8/12/16 channels + 2 digital marker lines) pairs with any supported DAQ; resolution, input range, and terminal mode are set by the DAQ you choose. BXI Studio handles acquisition for the devices below, and the list grows over time.
NI USB-6001
14-bit
±10 V
~1.2 kHz per channel across 4 channels
RSE or differential
Entry-level multifunction USB DAQ
NI mioDAQ (USB-6451)
20-bit
±0.2 / ±2.5 / ±5 / ±10 V
4–5 kHz per channel
RSE or NRSE
Reference device for this sheet
NI cDAQ (9189 + NI 9234)
24-bit
±5 V
Up to 10 kHz per channel
Pseudodifferential
Modular CompactDAQ; highest resolution
Terminal mode depends on the DAQ. Read the channels against the shared AGND in whichever mode your device offers, and set it explicitly rather than accepting the default.
Warning: The mioDAQ powers on in differential mode at the ±10 V range, which pairs AI 0 with AI 8 and reads railed noise from a single-wire hookup. Select RSE before acquiring.
Note: cDAQ terminal modes are set by the module rather than the chassis. The NI 9234 used here offers pseudodifferential only, and other C Series modules differ. Check your module's specifications.
Note: Sampling rate belongs to the DAQ, not to xBud. On multiplexing DAQs the maximum is an aggregate figure shared across the channels in the task, so the rate available per channel falls as you add channels. Size the DAQ to the channel count and bandwidth your experiment needs.
The table below lists the NI mioDAQ as the reference example.
Electrical: reference DAQ (NI mioDAQ USB-6451)
Model
NI USB-6451
ADC resolution
20-bit
Input voltage range
±0.2 / ±2.5 / ±5 / ±10 V (software-selectable)
Analog input channels
16 single-ended / 8 differential
Sampling rate
Up to 1 MS/s aggregate; about 4–5 kHz per channel in typical xBud use
Digital I/O
Digital inputs accept the two marker lines
Connectivity / power
USB-C, bus-powered
Note: These are the values the xBud workflow relies on. For the complete and authoritative mioDAQ specifications and compliance declarations, see the NI USB-6451 specifications.
System (end-to-end)
Recommended acquisition rate (EEG)
≧ 1000 Hz
User-configurable up to the DAQ maximum, see Configure signal settings
End-to-end input-referred noise
< 10 µV
xBud + DAQ, 0.5–45 Hz
Effective resolution at the electrode
See the table below
Set by the DAQ's resolution and input range, divided by the fixed 100× gain
System bandwidth (−3 dB)
0.5 - 300 Hz
Set by the xBud filters
Resolution at the electrode is not a property of xBud. It follows from the DAQ you chose, the range you set it to, and the fixed 100× gain, so it is to be read with those conditions attached.
95 nV
20-bit
±5 V
191 nV
20-bit
±10 V
1.53 µV
16-bit
±5 V
3.05 µV
16-bit
±10 V
Note: The reference mioDAQ is 20-bit. The 16-bit rows are there because the same xBud front-end pairs with lower-resolution DAQs, and the figure changes by more than thirty-fold across this table. A resolution quoted without its DAQ and range says nothing.
Tip: Set the DAQ to the narrowest input range the signal allows. The fixed 100× gain puts a 100 µV EEG signal at about 10 mV, which uses roughly 0.1% of a ±10 V range. Selecting ±2.5 V instead recovers two bits of effective resolution. Surface EMG at a few millivolts is comfortable on any range.
To set the input range and the sampling rate in BXI Studio, see Configure signal settings.
Mechanical and physical
xBud
Dimensions (L × W × H)
110 × 170 × 30 mm (4-channel; larger variants to be confirmed)
Weight
300 g
Enclosure material / finish
Metal
Electrode connector
Standard EXG electrode connectors for dry Ag/AgCl electrodes (comb contacts through hair, disc contacts on bare skin)
Analog output connector
Standard jumper-wire header to the mioDAQ terminals
Indicators
Three LEDs: two for digital markers, one for system power
Controls
None. Gain and bandwidth are fixed in hardware
xBud signal map
EXG channel outputs
4 / 8 / 12 / 16
Conditioned biopotential, post-gain
AI0…AI(n−1), read against AGND
AGND
1
Analog ground, shared between xBud and mioDAQ
AI GND
REF
1
Recording reference electrode on the subject (for example, Cz for EEG); used within xBud conditioning
Optionally digitised
DRL
1
Driven-right-leg electrode for common-mode noise cancellation
Subject only (not digitised)
Marker lines
2
Driven by the two-button marker box, which cables to the xBud chassis
DAQ digital inputs (e.g. port0/line0–1)
Note: xBud amplifies every electrode, including REF, differentially against AGND, then amplifies the EEG electrodes against REF. The common-mode signal drives the DRL electrode for noise cancellation. Read the channels at the DAQ against the shared AGND, in the terminal mode your DAQ offers: RSE on the mioDAQ, pseudodifferential on a cDAQ with the NI 9234.
Note: Markers can be triggered in hardware (the two-button marker box) or in software (a BXI Studio keypress).
To set up markers in BXI Studio, see Marker configurations.
NI mioDAQ (USB-6451)
Form factor
Terminal block with spring terminals and back-shell strain relief
Weight
598 g
Connector to host
USB-C (locking, 2 m cable supplied)
Terminal access
Spring terminals; NI screwdriver supplied
Dimensions
117 x 177 x 31 mm (from the NI USB-6451 datasheet)
Power
xBud power adapter
2-pin floating AC-DC adapter; 110–220 V AC in, 12 V DC / 2 A out, barrel jack
Use only the supplied adapter
xBud power consumption
~5 W
About 300 mA at ±5 V for the 4-channel, plus overhead
DAQ power
Set by the DAQ
the NI mioDAQ is USB-C bus-powered
Warning: Separation from the mains comes from the supplied adapter, not from xBud. Power xBud only from that adapter. It is not certified to IEC 60601-1 and is not characterised for patient leakage current. It is a teaching aid, not a medical device. Read Power & Safety before connecting any subject.
Environmental
Operating temperature
0 to 40 °C
Operating humidity
20 to 80 % RH, non-condensing
Storage temperature
-20 to 60 °C
Compliance
RoHS
Note: These are the conditions xBud is intended to be used and stored in, which is an ordinary laboratory or classroom. The components inside tolerate a wider range, but the assembled product is not environmentally tested, so the wider figures are not published as a rating.
Regulatory and safety
Intended use
Research, education, and prototyping only, not a medical device
Diagnostic use
Prohibited
Electrical isolation
From the supplied AC-DC adapter only; xBud adds no barrier. Not certified to IEC 60601-1. See Power & Safety
Certifications and compliance
See the regulatory portal link below
Ordering information
xBud 4-channel
4 EXG + REF/AGND/DRL
Entry: ECG, EMG, single-region EEG
xBud 8-channel
8 EXG + REF/AGND/DRL
Intermediate multimodal
xBud 12-channel
12 EXG + REF/AGND/DRL
Multi-region EEG
xBud 16-channel
16 EXG + REF/AGND/DRL
Advanced BCI and dense montage
Subject interfaces
Order any or all of the four interfaces with any channel variant. The interface determines where electrodes sit on the body, not what the front-end does.
EEG cap
EEG, full montage
Comb contacts through hair, disc contacts on bare skin
EEG headband
EEG, frontal or occipital
Comb and disc contacts
EMG armband
Surface EMG
Disc contacts
ECG chest-strap
ECG
Disc contacts
What's in the box
xBud AFE in the selected channel variant
Standard Ag/AgCl coated dry electrode set (comb and disc contacts) and lead cables with gold-plated adaptors
The subject interfaces ordered: EEG cap, EEG headband, EMG armband, ECG chest-strap
xBud line-power adapter
Marker box (two buttons with LEDs) and connecting cable
A supported DAQ (NI USB-600x, NI mioDAQ, or NI cDAQ) with its cable and terminal accessories
Jumper wires for the xBud-to-DAQ connection
BXI Studio software access
Software
BXI Studio
Primary acquisition and visualisation
NI-DAQmx driver
Required runtime; install before BXI Studio can detect the mioDAQ
NI-DAQmx APIs (Python, C/C++)
Programmatic access to the mioDAQ
See SDK & Dev Resources for code examples.