Configure electrodes

This is the final Instinct configuration screen. This is where you fine-tune each electrode's position and verify signal quality through real-time impedance monitoring.

Screen layout

Left panel: EEG diagram

An interactive head diagram shows all 15 electrode positions. The static reference layout matches the channel map shipped with the Instinct headset:

Each electrode is color-coded by impedance:

Color
Impedance
Meaning

Green

< 50 kΩ

Good contact: ready for recording

Yellow

50–200 kΩ

Moderate: may work, but adjusting is recommended

Red

> 200 kΩ

Poor contact: needs adjustment

Zoom and pan the diagram with the controls in the top-right corner:

  • +: zoom in

  • −: zoom out

  • Fit: reset to fit the full diagram

Click an electrode on the diagram to select it.

Right panel: electrode list

A scrollable list of all 15 electrodes. Each row shows:

Element
Description

Name

Electrode identifier (FP1, FP2, F3, F4, C3, C4, P3, P4, O1, O2, F7, F8, T3, T4, CZ)

Position

Current motor position in mm (0–20 mm range)

Reset

Refresh icon to reset the electrode to position 0

Down/Up

Motor control arrows (click or long-press)

Impedance

Real-time impedance reading in kΩ

Status bar

Color-coded right border matching impedance quality

Adjust an electrode

  1. Click a row or click on the diagram to select an electrode.

  2. The selected row highlights in purple.

  3. Use the Up/Down arrows to move the motor:

    • Click: move one step

    • Long press: continuous movement (hold the button)

  4. Watch the impedance value update in real time.

  5. Goal: get all electrodes to green (< 50 kΩ).

Tips for good impedance

  • Move electrodes down to increase scalp contact.

  • If impedance stays red, the cause is usually hair blocking the electrode.

  • Ask the subject to part their hair at that electrode location.

  • Use the reset button to return an electrode to position 0 and start over.

Options menu

Click the three-dot menu (top-right) for additional options:

  • Reset All Electrodes: move all motors back to position 0.

  • Recalibrate: return to the Calibration screen.

  • Disconnect: disconnect the device.

Complete configuration

  1. Adjust electrodes until most show green impedance.

  2. Click Continue at the bottom-right.

  3. The Configuration Complete modal appears:

    • Enter or confirm the subject name.

    • Check Save subject and configurations to persist motor positions.

    • Click Finish.

  4. BXI Studio navigates you to the Dashboard.

Tip: Saved configurations mean the next time this subject uses this device, the electrodes automatically move to their saved positions.

Continue to set gain and frequency

Click Next to proceed to set gain and frequency, where you configure the channels gains and low pass frequency values.