Motor Calibration

Applies to: Instinct EEG Headset

The Instinct headset has 15 motorized electrodes that physically extend and retract to make contact with the scalp. Calibration zero-references each motor's position. Without calibration, electrode positioning may be off, leading to poor signal quality.

When motor calibration is required

Scenario
Required?

First time using a device

Mandatory: cannot skip

Within 30 days of last calibration

Optional: reminder appears, can continue without

More than 30 days since last calibration

Mandatory: must calibrate before proceeding

After firmware update

Recommended

The calibration warning modal

When you enter the calibration screen, a modal appears automatically.

Optional motor calibration (within 30 days)

  • Shows when the device was last calibrated and how many days remain.

  • Recalibrate: dismiss modal and proceed to calibrate.

  • Continue: skip calibration entirely, go to Subject Selection.

  • Optional "Do not remind me again" checkbox.

Mandatory motor calibration (overdue)

  • Shows "This device needs to be calibrated now".

  • Only one button: Calibrate: dismisses the modal.

  • Cannot skip: the Skip This Step button is hidden.

Run a motor calibration

  1. Place the headset on a flat surface. This is critical: do not wear it during calibration.

  2. Set the number of calibration cycles:

    • Default: 2 cycles

    • Range: 1–10 cycles

    • More cycles = more precision, but takes longer

    • Use the + and − buttons to adjust

  3. Click Calibrate to start.

  4. A progress bar shows real-time calibration status.

  5. Each cycle takes 15–30 seconds.

  6. When complete, click Done to proceed to Subject Selection.

What happens during motor calibration

The device runs each of its 15 motors through their full range of motion (0–20 mm), recording the exact positions. The app stores this position map and uses it for subsequent electrode adjustments.

Skip calibration

If calibration is optional, click Skip This Step at the bottom-left to go directly to Subject Selection. Not recommended for first-time use or if you notice inaccurate electrode positioning.

Next steps

  • Set up a Subject: Start configuration for connected device (applicable to Instinct)